When cleaning surgical instruments, hot water is often considered helpful for removing soils and improving detergent performance. However, higher water temperatures do not necessarily mean better cleaning, particularly during the prewash phase of an automated washer-disinfector.
Using excessively hot water at this stage can cause protein-based soils, such as blood, to coagulate and adhere more firmly to instrument surfaces, making subsequent cleaning more difficult.
How does this happen? And why must water temperature be carefully controlled during prewash?
1. Why Can Excessively High Water Temperatures Affect Instrument Cleaning Performance?
After use, surgical instruments often carry residual blood, tissue, and other organic soils. Many of these materials contain proteins whose structures are sensitive to temperature.
When exposed to elevated temperatures, proteins can undergo denaturation, which alters their original molecular structure. Depending on the protein and the exposure conditions, this process can lead to coagulation, causing protein-based soils to adhere more firmly to instrument surfaces.
An everyday example is cooking an egg white. Heat causes its proteins to denature, transforming the egg white from a transparent liquid into an opaque, solid substance.
A similar process can occur when blood and other protein-based soils on surgical instruments are exposed to excessively high temperatures. Once these proteins coagulate and become more firmly attached, they can be harder to remove during subsequent cleaning stages.
This is why water temperature needs to be appropriately controlled from the very beginning of an automated washing and disinfection cycle.
2. The Role of Water Temperature During Prewash
In an automated washer-disinfector cycle, the prewash phase removes loosely attached soils from instrument surfaces before the main wash begins.
Using water at a suitably low temperature during this phase helps flush away blood and other organic soils while limiting the risk of heat-induced protein coagulation.
Conversely, if excessively hot water is introduced before protein-based soils have been adequately removed, these soils may adhere more firmly to instrument surfaces, making subsequent cleaning more difficult.
The key is to select the appropriate temperature for each stage of the washing and disinfection cycle, rather than applying high temperatures from the outset.
3. What Does ISO 15883-1 Specify About Prewash Water Temperature?
According to SteelcoBelimed’s reference to ISO 15883-1—the international standard for washer-disinfectors—water supplied during the prewash phase must be sufficiently cool to prevent protein coagulation.
In particular, the standard notes that:
-
Temperatures above 45°C can cause protein coagulation, leading to difficulties in instrument cleaning. However, 45°C is not a universal target temperature for all prewash cycles.
-
Depending on the characteristics of the instruments and the validated washing and disinfection cycle parameters, the required water temperature may be considerably lower.
In general, using cooler water during prewash can support the removal of protein-based soils. However, actual operating temperatures must follow the manufacturer’s instructions and validated cycle parameters.
4. Temperature Control – A Key Factor in Automated Washing and Disinfection
The effectiveness of an automated washing and disinfection cycle depends on the combined action of four main factors:
-
Temperature: Controlled to meet the requirements of each processing stage.
-
Mechanical action: Helps remove soils from instrument surfaces.
-
Cleaning chemicals: Help break down and remove soils.
-
Time: Ensures that processing stages follow the established parameters.
The prewash phase illustrates the importance of controlling both the sequence of processing stages and their operating conditions.
If excessively high temperatures are introduced before protein-based soils have been adequately removed, cleaning performance in subsequent stages may be compromised—even when those stages operate within the specified parameters.
Establishing and maintaining the appropriate temperature at each stage is therefore an important requirement for an effective automated washing and disinfection cycle.
5. Conclusion
Higher water temperatures do not necessarily deliver better surgical instrument cleaning.
During prewash in particular, appropriate water temperature control helps limit protein coagulation and supports the removal of blood and other organic soils before the main wash.
An effective washing and disinfection cycle requires careful coordination of temperature, time, mechanical action, and cleaning chemicals. Each stage must operate under appropriate, validated parameters.
Controlling temperature correctly from the prewash phase is an important factor in improving cleaning performance and the quality of medical instrument reprocessing.
DUONG GIA – MEDICAL INSTRUMENT CLEANING AND DISINFECTION SOLUTIONS
As an authorized distributor of SteelcoBelimed in Vietnam, Duong Gia provides automated washer-disinfector solutions, supporting healthcare facilities in improving cleaning and disinfection performance and optimizing instrument reprocessing workflows.
For more information or specialist advice on automated washer-disinfector systems, please contact Duong Gia at 0889 930 615.
Reference: SteelcoBelimed – “Washer-Disinfectors: Why Temperature Matters During Prewash.”
Author: Jonathan Carfora – Technical Services Trainer.
Read the original article on the SteelcoBelimed website.